OLYMPUS CF-H190L Electronic Colonoscope For Repair
► Brief Description
Inspection Results:
1. Image display present;
2. Partial broken filament in the light bundle, two bundle glasses worn;
3. Front end of insertion tube damaged;
4. Upper angle malfunctioning, other angles stiff;
5. Light guide tube damaged;
6. C-cover worn;
7. Leak at No. 1 button.
Maintenance plan:
Solution1: Replace C-cover, two bundle glasses, insertion tube, snake bone, steel wire, helix tube, light guide tube, No. 1 button, and rubber . CCD not covered under warranty.
Solution 2: Replace snake bone, steel wire, helix tube, No.1 buttons and rubber.
► Common issues with OLYMPUS CF-H190L
Mechanical & Structural Faults
Image System Faults
Leakage & Air/Water System Faults
Light Guide System Faults
►FAQ
Insertion tube has serpentine movement and outer rubber cracking. What parts need replacement?
A: Full insertion tube, helix tube and outer rubber assembly must be replaced to restore original mechanical performance.